Evaluation of the Etiologies of Implant Fracture in Patients With Fractures of the Implants of Lower Limbs Long BonesAli Yeganeh, Babak Otoukesh, Peyman Kaghazian, Nima Yeganeh, Bahram Boddohi, Mehdi Moghtadaei. Abstract | | | | Background: Orthopedics implants are important tools for treatment of bone fractures. Despite available recommendations for designing and making the implants, there are multiple cases of fracture of these implants in the body. Hence, in this study the frequency of failure of implants in long bones of lower extremities was evaluated. Methods and Materials: In this cross-sectional study, two types of fractured implants in the body were analyzed and underwent metalogical, mechanical, and modeling and stress-bending analysis. Results: The results revealed that the main cause of fractures was decreased mechanical resistance due to inappropriate chemical composition (especially decreased percentages of Nickel and Molybdenum). Conclusions: It may be concluded that following the standard chemical composition and use of optimal making method are the most important works for prevention of failure of implants.
Key words: Implant, SEM, Optical Metallography, Quantometry.
|
|
|
|